Observation Possible Cause Action
Excess template was loaded onto
the chip.
In the QuantStudio
®
3D
AnalysisSuite
Software, check
the copies per reaction and dilute
the sample accordingly so that it
is within the digital range.
Chapter 5 Troubleshooting
Troubleshooting chip images using the Chip View
5
84
QuantStudio
®
3D Digital PCR System User Guide